Measurement Principle
Ion sensors allow flexible selection of one or two ion-selective electrodes, while standard configurations include pH and temperature electrodes by default, enabling simultaneous monitoring of four parameters. Optional ion-selective electrode parameters include: ammonium ion (NH4+), nitrate ion (NO3-), potassium ion (K+), chloride ion (Cl-), and fluoride ion (F-). Different parameter combinations can also provide mutual compensation, resulting in more accurate measurement data.
Features
lOne or two electrodes can be optionally selected from ammonium, nitrate, potassium, chloride, and fluoride ions. Flexible combinations are available to meet different customer needs.
lDigital sensor with strong anti-interference capability and long transmission distance;
lStandard digital signal output enables integration and networking with other devices without a transmitter;
lEasy and quick on-site installation of the sensor, supporting plug-and-play;
lReverse polarity protection for sensor power supply; protection against incorrect connection of RS485 A/B terminals to power;
lFeatures time display, data storage, and historical data review;
lOptional wireless data transmission module.
Typical Applications
lMonitoring ammonia nitrogen, nitrate nitrogen, and other measurements in surface water, groundwater, industrial processes, and wastewater treatment.
Sensor Technical Parameters
Measurement Range | Ammonia Nitrogen (NH4-N): 0.1~1000 mg/L Nitrate Nitrogen (NO3-N): 0.5~1000 mg/L Potassium Ion (K+): 0.5~1000 mg/L Chloride Ion (Cl-): 3~1000 mg/L Fluoride Ion (F-): 0.5~1000 mg/L pH: 5-10 Temperature: 0-40°C |
Resolution | Ion Parameters: 0.01 mg/L Temperature: 0.1°C pH: 0.01 |
Measurement Accuracy | Ammonia Nitrogen: ±5% of reading or ±0.2 mg/L, whichever is greater Other Ion Parameters: ±10% of reading or ±0.2 mg/L, whichever is greater Temperature: ±0.5°C pH: ±0.1 pH |
Repeatability | ±5% or 0.2 mg/L, whichever is greater |
Communication Protocol | MODBUS RS485 |
Storage Temperature | 0 to 50°C (non-freezing) |
Operating Temperature | 0 to 40°C (non-freezing) |
Dimensions | 55mm × 342mm (Diameter × Length) |
Weight | <1 KG |
Protection Rating | IP68/NEMA6P |
Cable Length | Standard 10-meter cable, extendable up to 100 meters |
DS-115T Controller Technical Parameters:
Display Output | 128*64 graphic dot-matrix LCD with strong LED backlight, operable under direct sunlight |
Power Supply | AC Power: 85-500 VAC (50/60Hz); DC Power: 9~36 VDC |
Main Materials | PA66+GF25+FR (upper cover); powder-coated aluminum alloy (lower shell) |
Output | Three channels of 4-20mA |
Relay | Three relays, with programmable response parameters and values |
Digital Communication | Equipped with MODBUS RS485 communication function for real-time transmission of measured values |
Storage Temperature | -20 to 70°C |
Operating Temperature | -15 to 60°C |
Waterproof Rating | IP65/NEMA4X |
Dimensions | 145mm*125mm*162mm (L*W*H) |
Weight | 1.35 KG |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-ISE900 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |